The 20 MeV version of MIRRORCLE produces a brilliance, as defined by photon flux per unit solid angle and area, of about 10
14 (photons/sec/mrad
2/mm
2/0.1%λ) as shown in table 1 and figure 1. The high brilliance and energy tunability makes laboratory based advanced analysis a reality. There are two methods to produce monochromatic X-rays, with the selected method based on experimental requirements.
Method A is to generate low energy bremsstrahlung radiation from carbon or beryllium target. Then the bremmsstrahlung radiation is monochromatized by one crystal. Method B is to generate monochromatic X-ray directly from crystal target. Crystal target and downstream crystal completes as a double crystal monochromator. The X-ray energy is tunable with fixed sample position using either method.
/technology/th_Brilliance-e.gif)
Figure 1. The calculated brilliance in units of photons/sec/mrad
2/mm
2/0.1%λ. The output spectrum of 20 MeV model is compared to 6 MeV model for carbon target. Using the 20 MeV model, a laboratory size protein crystallography beamline can be constructed.
(A) White bremsstrahlung X-rays are monochromatized by a single crystal. The 20 MeV model produces X-rays with a deviation of ±1.5°. By rotating the crystal the final X-ray energy output is changed and resulting mechanism is can be considered part of a low cost beam line.
The energy resolution of the monochromatic beamline is E/E = 1000 with 0.1 mrad aperture. A high energy X-ray may form background when bremsstrahlung radiation is used.
(B) A small crystal placed in the electron orbit generates directly monochromatic X-rays (Parametric X-ray radiation), and the observed spectrum change is shown in figure below. The X-ray energy is tunable with rotation of crystal target. Combining this crystal target with a second crystal, a double crystal monochromator beamline is formed. It takes about 1 second to change the X-ray energy. We can obtain monochromatic X-ray in the range of 8-30keV. The background is very low, because light axis of parametric X-ray radiation is different from the axis of bremsstrahlung radiation.
Figure 2. The beamline with method (B).
Figure 3. Monochromatic X-rays from Si (111) crystal target using MIRRORCLE-20SX with method (B). "Rotation angle of crystal" means the tangential direction of the beam orbit. The graph shows energy tunability rotating crystal target.
Detector : Si PIN photodiode.
Figure 4 shows the second upright monochromator chamber for MIRRORCLE-20SX. This system is located in the Omi MIRRORCLE Center in Ritto-city, Japan. MIRRORCLE-20SX was successfully commissioned, and is producing bright X-rays. There is a CCD detector downstream of a goniometer for protein structure analysis. The sample positon for protein crystal is 3.5 m distance from X-ray source point. The short beamline results in high intensity X-ray at the sample location.
Fig4. Photo of the second monochromator chamber
Success in the measurement of MIRRORCLE-20SX absolute X-ray intensity.
We have measured the monochromatized X-ray intensity through a sagittal
monochrometer for 10-15 keV from MIRRORCLE-20SX. An imaging plate was used to compare the data
from the X-ray tube. The same sagittal monochrometer was used for 900W Cu tube (8keV).
In terms of brilliance, it corresponds to about 10
12 photons. Of course the difference between
MIRRORCLE and any X-ray tube is the energy scanning capability of MIRRORCLE.
We found that the measurement of the DXAFS spectrum can be made in 10 minutes.
For the first time a laboratory sized XAFS or protein crystallography system is now possible. In the
near future we will increase the X-ray flux by more than 100 times by implementing a new accelerator cavity.
(Note: X-ray tube needs more power to produce 13.4keV X-ray)
Table 1. A comparison X-ray intensity between MIRRORCLE and X-ray tube.
|
MIRRORCLE RAY
20 MeV | MIRRORCLE RAY
4 MeV | X-ray tube
| Synchrotron Radiation Facility
(8GeV Bending Magnet) |
| Figure of Radiation Mechanism |
/monochro/small_MIRRORCLE_radiation.png) |
/monochro/small_MIRRORCLE_radiation.png) |
/monochro/small_microfocus.png) |
/monochro/small_Synchrotron.png) |
| Source size | 10 μm φ | 10 μm φ | 3mmφ (The case of Micro-focus; 10μmφ) | 1 mm x 20 μm |
| Density @13keV (photons/sec/mrad2/0.1%λ) |
1010 (Present Value: 108) |
108 (Present Value: 108) |
108 @8keV |
1011 |
| Brilliance@13keV (photons/sec/mrad2/mm2/0.1%λ) |
1014 (Present Value: 1012) |
1012 (Present Value: 1012) |
(The case of Micro-focus; 1012 @8keV) |
1015 |
| Total Flux from X-ray port |
10kW (Present Value: 100W) |
15W
|
About 1 W |
170 W |
Dispersive XAFS beamline
Dispersive XAFS (DXAFS) can be conducted using monochromatized white bremsstrahlung X-rays.
MIRRORCLE RAY 20 MeV model is inherently suitable for DXAFS owing to its micrometre-size
X-ray source point and wide radiation emission angle. DXAFS experiments can be performed
with high energy resolution and over a wide range.
/XAFS/th_XAFS.png)
Fig5. The absorbance graph is changed from CCD X-ray image. The absorption edge or XAFS beats are observed. The XAFS sample is Mo foil (10 μm thickness), and we have used Si (220) monochromator.
Table 2. The product for Monochromatic X-ray user
MIRRORCLE type:
|
20 MeV model (20ST) |
MIRRORCLE RAY 4 MeV |
X-ray emission scheme
|
Bremsstrahlung
(e.g. 1 x 1 x 0.5 mm carbon target) |
Parametric X-ray
(e.g. 1 x 6 x 0.05 mm Si (111) target) |
Bremsstrahlung
(e.g. 1 x 1 x 0.5 mm carbon target) |
Emitted X-ray
|
10 - 20000 keV (White X-ray) |
8 - 30 keV (Monochromatic X-ray) |
10 - 4000 keV (White X-ray) |
| Irradiation area |
25 mrad Φ |
125 mrad Φ |
| Maximum photon power |
10 kW |
15 W |
Density (Brightness)
[photons/s/mrad2/0.1% λ] |
1010 |
108 |
Brilliance
[photons/s/mrad2/mm2/0.1% λ] |
1014 |
1012 |
| Maximum power input |
200 kVA |
50 kVA |
| Total size |
W 210 x D 450 x H 143 cm |
W 60 x D 150 x H 60 cm |
Total weight
|
3.75 t |
1 t(with a Pb shield) |
*1 MIRRORCLE's X-ray power very much depends on the vacuum level and is expected to improve over time.
Conveniently printable specifications with all models compared can be
found in the
Literature & CD section
of this web site.
Due to continuous improvement, features, specifications and price are subject to change without notification.
Reference:
- H. Yamada, AIP Conf. Proc. 902, pp.11-18, (2007).
- H. Yamada, T. Hanashima, M. Morita, and Y. Yano, AIP Conf. Proc. 902, pp.26-30, (2007).
- H. Yamada, H. Saisho, T. Hirai and J. Hirano, Spectrochimica Acta B 59, pp. 1323-1328, (2004).
Laue pattern taken by MIRRORCLE-6X:
/Laue/th_LauePattern.png)
We show the measured Laue pattern from a sample SiO
2 crystal (Quartz) using white X-ray generated by common target using MIRRORCLE-6X.
X-ray Fluorescence Analysis:
MIRRORCLE technology can provide the X-ray levels necessary to perform X-ray fluorescence analysis
of heavy elements. These examples show the spectrums for lead and platinum.